Daq rates, Table 41. analog input settling time versus gain, Table 42. pci1200 maximum recommended daq rates – National Instruments PCI-1200 User Manual

Page 62: Daq rates -10, Table 4-1, Analog input settling time versus gain -10, Table 4-2, Pci-1200 maximum recommended daq rates -10

Advertising
background image

Chapter 4

Theory of Operation

PCI-1200 User Manual

4-10

© National Instruments Corporation

DAQ Rates

Maximum DAQ rates (number of samples per second) are determined by
the conversion period of the ADC plus the sample-and-hold acquisition
time. During multichannel scanning, the DAQ rates are further limited by
the settling time of the input multiplexers and programmable gain
amplifier. After the input multiplexers are switched, the amplifier must be
allowed to settle to the new input signal value to within 12-bit accuracy
before you perform an A/D conversion, or 12-bit accuracy will not be
achieved. The settling time is a function of the gain selected.

Table 4-1 shows the recommended settling time for each gain setting
during multichannel scanning. Table 4-2 shows the maximum
recommended DAQ rates for both single-channel and multichannel data
acquisition. For single-channel scanning, this rate is limited only by the
ADC conversion period plus the sample-and-hold acquisition time,
specified at 10

µs. For multichannel data acquisition, observing the DAQ

rates in Table 4-2 ensures 12-bit resolution. The hardware is capable of
multiple scanning at higher rates than those listed in Table 4-2, but 12-bit
resolution is not guaranteed.

Table 4-1. Analog Input Settling Time Versus Gain

Gain

Settling Time

(Accuracy ±0.024% [±1 LSB])

1

10 µs typ, 14 µs max

2–10

13 µs typ, 16 µs max

20

15 µs typ, 19 µs max

50

27 µs typ, 34 µs max

100

60 µs typ, 80 µs max

Table 4-2. PCI-1200 Maximum Recommended DAQ Rates

Acquisition Mode

Gain

Rate

Single-channel

1, 2, 5, 10, 20, 50, 100

100 kS/s

Multichannel

1

2, 5, 10

20
50

100

100 kS/s

77 kS/s

66.6 kS/s

37 kS/s

16.6 kS/s

Advertising