Selecting trigger timer arm source – VXI VT1422A User Manual

Page 132

Advertising
background image

130 Programming the VT1422A for Data Acquisition and Control

Chapter 4

NOTES

1. When TRIGger:SOURce is not TIMer, ARM:SOURce must be set to

IMMediate (the *RST condition). If not, the INIT command will
generate an error -221,"Settings conflict."

2. When TRIGger:SOURce is TIMer, the trigger timer interval

(TRIG:TIM <interval>) must allow enough time to scan all channels,
execute all algorithms and update all outputs or a +3012, "Trigger
Too Fast" error will be generated during the trigger cycle. See the
TRIG:TIM command on page 397 for details.

To set the trigger source to the internal Trigger Timer (the default):

TRIG:SOUR TIMER

now select ARM:SOUR

To set the trigger source to the External Trigger input connection:

TRIG:SOUR EXT

an external trigger signal

To set the trigger source to a VXIbus TTLTRG line:

TRIG:SOUR TTLTRG1

the TTLTRG1 trigger line

Selecting Trigger Timer

Arm Source

Figure 4-6 shows that when the TRIG:SOUR is TIMer, the other trigger
sources become Arm sources that control when the timer will start.
The command to select the arm source is ARM:SOURce <source>.

The <source> parameter choices are explained in the following table.

NOTE

When TRIGger:SOURce is not TIMer, ARM:SOURce must be set to
IMMediate (the *RST condition). If not, the INIT command will generate
an error -221,"Settings conflict."

To set the external trigger signal as the arm source:

ARM:SOUR EXT

trigger input on
connector module

Parameter Value

Source of Arm (after INITiate:

command)

BUS

ARM[:IMMediate]

EXTernal

"TRG" signal input on terminal module

HOLD

ARM[:IMMediate]

IMMediate

The arm signal is always true (scan starts when an
INITiate:

command is received).

SCP

SCP Trigger Bus (future SCP Breadboard)

TTLTrg<n>

The VXIbus TTLTRG lines (n=0 through 7)

Advertising