VT1422A Command Reference 369

Chapter 7

[SENSe]

[SENSe:]STRain:UNSTrained

[SENSe:]STRain:UNSTrained <unstrained_v>,(@<ch_list>)

specifies the

unstrained voltage value to be used to convert strain bridge readings for the channels
specified by <ch_list>. The VT1529A/B can use the MEAS:VOLT:UNSTrained
command which automatically measures the unstrained bridge values and sends each
value to the channels’ EU conversion. This command does not control the output
voltage of any source.

Parameters

Comments

Use a voltage measurement of the unstrained strain bridge sense channel to

determine the correct value for <unstrained_v>.

The <ch_list> parameter must specify the channel used to sense the bridge

voltage, not the channel position on a Bridge Completion SCP.

Related Commands: SENSE:STRAIN:UNST?, SENSE:FUNC:STRAIN

…,

MEAS:VOLT:UNSTRained

Power-on Condition: Unstrained voltage is zero

*RST Condition: unaffected by *RST

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

STRAIN:UNST .024,(@100)

set unstrained voltage for channel 0

[SENSe:]STRain:UNSTrained?

[SENSe:]STRain:UNSTrained? (@<channel>)

returns the unstrained voltage

value currently used for EU conversion for the sense channel specified by <channel>.
This command does not make a measurement.

Parameters

Comments

Returned Value: Numeric value of unstrained voltage. The data type is float32.

The <channel> parameter must specify a single channel only.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

unstrained_v

numeric (float32)

-16 through +16

volts

ch_list

channel list (string)

100 - 15731

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 15731

none