Usage, Route:sequence:define, Parameters – VXI VT1422A User Manual

Page 327: Comments

Advertising
background image

VT1422A Command Reference 325

Chapter 7

ROUTe

Notice that since there are only 502 CVT elements available for up to 512

possible remote channel specifiers, these last 10 channels (15722-15731) must
not be sent to the CVT or an error will be generated. Since the default data
destination is to BOTH the FIFO and CVT (Data Destination 3), any reference
in ROUT:SEQ:DEF to the last 10 remote channels must force the data
destination to FIFO only.

*RST Condition: To supply the necessary time delay before Digital inputs

are read, the analog input (AIN) scan list contains two entries for channel 0 (100).
This minimum delay is maintained by replacing these default channels
as others are defined by algorithms or ROUT:SEQ:DEF. The three other lists
contain no channels.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

Define mix of Remote and on-board channels. First 128 Remote channels
default to FIFO and CVT, next 64 Remote channels are directed to FIFO only
and last eight On-board channels default to FIFO and CVT.

ROUT:SEQ:DEF (@10000:10931,2(11600:11731),124:131)

ROUTe:SEQuence:DEFine?

ROUTe:SEQuence:DEFine? <type>

returns the sequence of channels defined in

the scan list.

Parameters

Comments

The channel list contents and sequence are determined primarily by channel

references in the ROUT:SEQ:DEF command and in any algorithms currently
defined. The SENS:REF:CHANNELS and SENS:CHAN:SETTLING
commands also effect the scan list contents.

The <type> parameter selects which channel list will be queried:

"AIN" selects the Analog Input channel list (this is the Scan List).
"AOUT" selects the Analog Output channel list.
"DIN" selects the Digital Input channel list.
"DOUT" selects the Digital Output channel list.

"DEST" does not requesting the contents of a channel list type, rather it requests
the Data Destination number for each channel in the "AIN" channel list.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

type

(string)

AIN | AOUT | DEST | DIN | DOUT

none

Advertising