Programming the VT1422A & VT1529A/B for Remote Strain Measurement 91

Chapter 3

Figure 3-20. Performing Shunt Calibration

Set up the analog input Scan List to

measure un-shunted channels

Measure unstrained bridge output

voltages and send to EU routines

Send gage factors to

channel EU routines

SENSe:STRain:GFACtor

MEASure:VOLTage:UNSTrained?

ROUTe:SEQuence:DEFine (@

)

<ch_list>

Measure bridge excitation

voltages and get from FIFO

MEASure:VOLTage:EXCitation?/SENSe:DATA:FIFO?

Measure unstrained bridge output

voltages and get from FIFO

MEASure:VOLTage:UNSTrained?/SENSe:DATA:FIFO?

To convert readings to

Engineering Units of strain

SENSe:FUNCtion:STRain:<bridge_type>. Also sets
bridge config switches and turns on excitation V

Set up Trigger System to scan

strain bridge channel

TRIG:SOURce IMM

Initiate Trigger System

INITiate[:IMMediate]

Retrieve un-shunted strain

readings from FIFO

SENSe:DATA:FIFO?

Select INTernal or EXTernal

shunt resistor

OUTPut:SHUNt:SOURce

Set up the analog input Scan List to

measure single shunted channel

ROUTe:SEQuence:DEFine (@

)

<shunt_channel>

Turn on shunt resistor

to slected channel

OUTPut:SHUNt:STATe ON (@

)

<shunt_channel>

Initiate Trigger System

(measure shunted strain channel)

INITiate[:IMMediate]

Retrieve shunted strain

reading from FIFO

SENSe:DATA:FIFO?