Cls *dmc *emc *emc – VXI VT1422A User Manual

Page 402

Advertising
background image

400 VT1422A Command Reference

Chapter 7

IEEE-488.2 Common Command Reference

Note

If Open Transducer Detect (OTD) is enabled when *CAL? is executed, the module
will disable OTD, wait one minute to allow channels to settle, perform the calibration
and then re-enable OTD. If OTD is turned off by the program before executing
*CAL?, it should also wait one minute for settling.

*CLS

*CLS

Clear Status Command. The *CLS command clears all status event registers

(Standard Event Status Event Register, Standard Operation Status Event Register,
Questionable Data Event Register) and the instrument's error queue. This clears the
corresponding summary bits (bits 3, 5, and 7) in the Status Byte Register. *CLS does
not affect the enable bits in any of the status register groups. (The SCPI command
STATus:PRESet does clear the Operation Status Enable and Questionable Data
Enable registers.) *CLS disables the Operation Complete function (*OPC command)
and the Operation Complete Query function (*OPC? command).

Send with VXIplug&play Function: hpe1422_cmd(...)

*DMC

*DMC <name>,<cmd_data>

Define Macro Command. Assigns one or a sequence

of commands to a named macro.

The command sequence may be composed of SCPI and/or Common commands.

<name> may be the same as a SCPI command, but may not be the same as a Common
command. When a SCPI named macro is executed, the macro rather than the SCPI
command is executed. To regain the function of the SCPI command, execute the
*EMC 0 command.

<cmd_data> is sent as arbitrary block program data (see “Arbitrary Block Program
and Response Data” on page 233).

*EMC

*EMC <enable>

Enable Macro Command. When <enable> is non-zero, macros are

enabled. When <enable> is zero, macros are disabled.

*EMC?

*EMC?

Enable Macro query. Returns either 1 (macros are enabled) or 0 (macros

are disabled).

Advertising
This manual is related to the following products: