VT1422A Command Reference 367

Chapter 7

[SENSe]

[SENSe:]STRain:GFACtor

[SENSe:]STRain:GFACtor <gage_factor>,(@<ch_list>)

specifies the gage factor

to be used to convert strain bridge readings for the channels specified by <ch_list>.

Parameters

Comments

The <ch_list> parameter must specify the SCP or RSC channel used to sense

the bridge voltage, not the channel position on a Bridge Completion SCP.

Related Commands: SENSE:STRAIN:GFAC?, SENSE:FUNC:STRAIN

*RST Condition: Gage factor is 2

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:STRAIN:GFAC 3,(@100:107)

set gage factor for channels 0 through 7

SENS:STRAIN:GFAC 2.2,(@10000:10931)

set gage factor for VT1529A/B channels
0000 through 0931 (128 channels)

[SENSe:]STRain:GFACtor?

[SENSe:]STRain:GFACtor? (@<channel>)

returns the gage factor currently set for

the sense channel specified by <channel>.

Parameters

Comments

Returned Value: Numeric value of gage factor. The data type is float32.

The <channel> parameter must specify a single channel only.

Related Commands: STRAIN:GFACTOR

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

Usage

STRAIN:GFAC? (@107)

query gage factor for channel 7

enter statement here

returns the gage factor set by STR:GFAC

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

gage_factor

numeric (float32)

1 - 5

none

ch_list

channel list (string)

100 - 15731

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 15731

none