338 VT1422A Command Reference

Chapter 7

[SENSe]

[SENSe:]DATA:FIFO:RESet

[SENSe:]DATA:FIFO:RESet

clears the FIFO of values. The FIFO counter is reset

to 0.

Comments

When Accepted: Not while INITiated

Related Commands: SENSE:DATA:FIFO

*RST Condition: SENSE:DATA:FIFO:RESET

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENSE:DATA:FIFO:RESET

Clear the FIFO

[SENSe:]FREQuency:APERture

[SENSe:]FREQuency:APERture <gate_time>,<ch_list>

sets the gate time for

frequency measurement. The gate time is the time period that the SCP will allow for
counting signal transitions in order to calculate frequency.

Parameters

Comments

If the channels specified are on an SCP that doesn't support this function,

an error will be generated. See the SCP's User's Manual for its capabilities.

Related Commands: SENSe:FUNCtion:FREQuency

*RST Condition: 0.001 s

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:FREQ:APER .01,(@144)

set channel 44 aperture to 10 ms

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

gate_time

numeric (float32)

0.001 to 1 (0.001 resolution)

seconds

ch_list

string

100 - 163

none