136 Programming the VT1422A for Data Acquisition and Control

Chapter 4

Below is an example command sequence for Figure 4-8. It assumes that the
FIFO mode was set to BLOCK and that at least one algorithm is sending
values to the FIFO.

following loop reads number of values in FIFO while algorithms executing

loop while "measuring" bit is true

see STAT:OPER:COND bit 4

SENS:DATA:FIFO:COUNT?

query for count of values in
FIFO

input n_values here

if n_values >= 16384

Sets the minimum block size to
transfer

SENS:DATA:FIFO:PART? n_values

ask for n_values

input read_data here

Format depends on FORMat cmd

end if

end while loop

following checks for values remaining in FIFO after "measuring" false

SENS:DATA:FIFO:COUNT?

query for values still in FIFO

input n_values here

if n_values

if any values...

SENS:DATA:FIFO:PART? n_values

input read_data here

get remaining values from FIFO

end if

Begin Data Retrieval

Exit Data Retrieval

Algorithm Stopped?

Any Values in FIFO?

Enough Values

in FIFO?

Execute Final Transfer

Command

Execute Bulk Transfer

Command

DATA:FIFO:COUNT?

DATA:FIFO:PART? <n_values>

no

no

no

yes

yes

yes

STAT:OPER:COND?

(bit 4 "measuring")

Figure 4-8. Controlling Reading Count