344 VT1422A Command Reference

Chapter 7

[SENSe]

Usage

program must put table constants into array table_block

DIAG:CUST:PIEC 1,table_block,(@100:107)

send characterized thermocouple table
for use by channels 0-7

SENS:FUNC:CUST:TC N,.25,(@100:107)

link custom thermocouple EU with chs
0-7, use reference temperature
compensation for N type wire.

SENSE:REF RTD,92,(@120)

designate a channel to measure the
reference junction temperature

include these channels in a scan list (REF channel first)

INITiate then TRIGger module

[SENSe:]FUNCtion:FREQuency

[SENSe:]FUNCtion:FREQuency (@<ch_list>)

sets the SENSe function to

frequency for channels in <ch_list>. Also configures the channels specified as digital
inputs.

Parameters

Comments

If the channels specified are on an SCP that doesn't support this function,

an error will be generated. See the SCP's User's Manual for its capabilities.

Use the SENSe:FREQuency:APERture command to set the gate time for the

frequency measurement.

Related commands: SENS:FREQ:APER

*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital

SCP channels

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:FUNC:FREQ (@144)

set channel 44's to frequency

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

ch_list

string

100 - 163

none