116 Programming the VT1422A for Data Acquisition and Control

Chapter 4

Supplying a Fixed Reference Temperature

The [SENse:]REFerence:TEMPerature <degrees_c> command
immediately stores the temperature of a controlled temperature reference
junction panel in the Reference Temperature Register. The value is applied
to all subsequent thermocouple channel measurements until another
reference temperature value is specified or measured. There is no need to use
SENS:REF:CHANNELS.

To specify the temperature of a controlled temperature reference panel:

SENS:REF:TEMP 50

reference temp = 50

°

C

Now begin scan to measure thermocouples.

Linking Strain

Measurements

Strain measurements usually employ a Strain Completion and Excitation
SCP (VT1506A, VT1507A, VT1511A) or VT1529A/B Remote Strain
Conditioning Unit. To link channels to strain EU conversions send the
[SENSe:]FUNCtion:STRain:<bridge_type> [<range>,](@<ch_list>)

<bridge_type> is not a parameter but part of the command syntax. The

following table relates the command syntax to bridge type. See the
VT1506A, VT1507A, and VT1511A SCPs’ user’s manual for bridge
schematics and field wiring information.

* These choices are only available with the VT1529A/B.
** This choice is only available with VT1529A/B channels that have had a

user-supplied resistor installed.

The <ch_list> parameter specifies which sense SCP channel(s) to link

to the strain EU conversion, not the strain bridge completion SCP
channels. This parameter does not specify channels on the
VT1506A/07A Strain Bridge Completion SCPs. It can specify any of
the lower four channels of a VT1511A SCP since these channels are
the sense channels used to measure the SCPs four bridge completion
channels.

Command

Bridge Type

:FBENding

Full Bending Bridge

:FBPoisson

Full Bending Poisson Bridge

:FPOisson

Full Poisson Bridge

:HBENding

Half Bending Bridge

:HPOisson

Half Poisson Bridge

[:QUARter]

Quarter Bridge (default)

:Q120

*

Quarter using VT1529A/B’s

internal 120

Ω resistor

:Q350

*

Quarter using VT1529A/B’s

internal 350

Ω resistor

:User

**

Quarter using VT1529A/B’s
user supplied resistor value