Subsystem syntax, Arm[:immediate, Comments – VXI VT1422A User Manual

Page 259: Usage, Arm:source, Parameters, Arm[:immediate] arm:source

Advertising
background image

VT1422A Command Reference 257

Chapter 7

ARM

Subsystem Syntax

ARM

[:IMMediate]
:SOURce BUS | EXTernal | HOLD | IMMediate | SCP | TTLTrg<n>
:SOURce?

ARM[:IMMediate]

ARM[:IMMediate]

arms the trigger system when the module is set to the ARM:SOUR

BUS or ARM:SOUR HOLD mode.

Comments

Related Commands: ARM:SOURCE, TRIG:SOUR

*RST Condition: ARM:SOUR IMM

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

ARM:IMM

After INIT, system is ready for trigger event

ARM

Same as above (:IMM is optional)

ARM:SOURce

ARM:SOURce <arm_source>

configures the ARM system to respond to the

specified source.

Parameters

Comments

The following table explains the possible choices.

See note about ARM subsystem on page 256.

When TRIG:SOURCE is TIMER, an ARM event is required only to trigger the

first scan. After that the timer continues to run and the module goes to the
Waiting-For-Trigger State ready for the next Timer trigger. An ABORT
command will return the module to the Trigger Idle State after the current scan
is completed. See TRIG:SOURce for more detail.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

arm_source

discrete (string)

BUS | EXT | HOLD | IMM | SCP

| TTLTrg<n>

none

BUS

ARM[:IMMediate]

EXTernal

“TRG” signal on terminal module

HOLD

ARM[:IMMediate]

IMMediate

The arm signal is always true (continuous arming).

SCP

SCP Trigger Bus (future SCP Breadboard)

TTLTrg<n>

The VXIbus TTLTRG lines (n=0 through 7)

Advertising
This manual is related to the following products: