Usage, Arm:source – VXI VT1422A User Manual

Page 260

Advertising
background image

258 VT1422A Command Reference

Chapter 7

ARM

While ARM:SOUR is IMM, simply INITiate the trigger system to start a
measurement scan.

When Accepted: Before INIT only.

Related Commands: ARM:IMM, ARM:SOURCE?, INIT[:IMM],

TRIG:SOUR

*RST Condition: ARM:SOUR IMM

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

ARM:SOUR BUS

Arm with ARM command

ARM:SOUR TTLTRG3

Arm with VXIbus TTLTRG3 line

ARM:SOURce?

ARM:SOURce?

returns the current arm source configuration. See the ARM:SOUR

command for more response data information.

Returned Value: Discrete, one of BUS, HOLD, IMM, SCP, or TTLT0 through

TTLT7. The data type is string.

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

ARM:SOUR?

An enter statement return arm
source configuration

Advertising