264 VT1422A Command Reference

Chapter 7

CALibration

Comments

Related Commands: CAL:VAL:RES, CAL:STOR ADC

When Accepted: Not while INITiated

Send with VXIplug&play Function: hpe1422_cmd(...)

Command

Sequence

CAL:CONF:RES

connect reference resistor to Calibration
Bus

*OPC? or SYST:ERR?

must wait for CAL:CONF:RES to
complete

(now measure ref resistor with external DMM)

CAL:VAL:RES <measured value>

Send measured value to module

CAL:STORE ADC

Store cal constants in non-volatile
memory (used only at end of complete
cal sequence)

CALibration:CONFigure:VOLTage

CALibration:CONFigure:VOLTage <range>,<zero_fs>

connects the on-board

voltage reference to the Calibration Bus. A measurement of the source voltage can
be made with an external multimeter connected to the H Cal and L Cal terminals on
the Terminal Module or the V H, V L,

Ω H, and Ω L terminals on the Cal Bus

connector when not using a terminal module. The <range> parameter controls the
voltage level available when the <zero_fs> parameter is set to FSCale (full scale).

Parameters

Comments

The <range> parameter must be within

±5% of one of the five following values:

0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, 16 V dc. This parameter may be
specified in millivolts (mv).

The expected FSCALE output voltage of the calibration source will be

approximately 90% of the nominal value for each range, except the 16 V range
where the output is 10 V.

When Accepted: Not while INITiated

Related Commands: CAL:VAL:VOLT, STOR ADC

Send with VXIplug&play Function: hpe1422_cmd(...)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

range

numeric (float32)

see comments

volts

zero_fs

discrete (string)

ZERO | FSCale

none