268 VT1422A Command Reference

Chapter 7

CALibration

CALibration:SETup

CALibration:SETup

causes the Channel Calibration function to be performed for

every module channel with an analog SCP installed (input or output). The Channel
Calibration function calibrates the A/D Offset and the Gain/Offset for these analog
channels. This calibration is accomplished using internal calibration references. For
more information see *CAL? on page 399.

Note that the scope of the *CAL? and CAL:SETup commands is limited to the
VT1422A and the SCPs it contains. They do not calibrate Remote Signal
Conditioning Units like the VT1529A/B. CAL:REMote? must be used in addition to
*CAL?/CAL:SETup for RSCs.

Comments

CAL:SET performs the same operation as the *CAL? command except that,

since it is not a query command, it doesn't tie-up the driver waiting for response
data from the instrument. If there are multiple VT1422As in a system, a
CAL:SET operation can be started on each and then a CAL:SET? command can
be executed to complete the operation on each instrument.

Related Commands: CAL:SETup?, *CAL?

When Accepted: Not while INITiated

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

CAL:SET

start SCP Calibration on 1st VT1422A

:

start SCP Calibration on more VT1422As

CAL:SET

start SCP Calibration on last VT1422A

CAL:SET?

query for results from 1st VT1422A

:

query for results from more VT1422As

CAL:SET?

query for results from last VT1422A

CALibration:SETup?

CALibration:SETup?

Returns a value to indicate the success of the last CAL:SETup

or *CAL? operation. CAL:SETup? returns the value only after the CAL:SETup
operation is complete.

Comments

Returned Value:

The data type for this returned value is int16.

Value

Meaning

Further Action

0

Cal OK

None

-1

Cal Error

Query the Error Queue (SYST:ERR?)

See error message “3026” on page 454

Also run *TST?

-2

No results available

No *CAL? or CAL:SETUP done

Advertising