Command sequence, Calibration:value:voltage, Parameters – VXI VT1422A User Manual

Page 276: Comments

Advertising
background image

274 VT1422A Command Reference

Chapter 7

CALibration

<ref_ohms> must be within 5% of the 7500

Ω nominal reference resistor value

or a -222 ’Data out of range’ error will be generated. If this error occurs,
verify the external measurement equipment and run *TST? on the VT1422A.

<ref_ohms> may be specified in kOhm (kohm).

When Accepted: Not while INITiated

Related Commands: CAL:CONF:RES, CAL:STORE ADC

Send with VXIplug&play Function: hpe1422_cmd(...)

Command

Sequence

CAL:CONF:RES

(now measure ref resistor with external DMM)

CAL:VAL:RES <measured value>

Send measured value to module

CALibration:VALue:VOLTage

CALibration:VALue:VOLTage <ref_volts>

sends the value of the just-measured dc

reference source to the module for A/D calibration.

Parameters

Comments

The "expected" output values for the voltage reference source is:

0.9 * Nominal Range Value for the 0.0625 through 4 volt ranges.
10 volts for the 16 volt range.

Use the CAL:CONF:VOLT command to configure the on-board voltage source

for measurement by an external reference voltmeter via the Calibration Bus
terminals.

A measurement of the source voltage can be made with an external multimeter

connected to the HCAL and LCAL terminals on the Terminal Module or the
V H and V L terminals on the Cal Bus connector when not using a terminal
module.

The <ref_volts> value given must be for the currently configured range and

output (zero or full scale) as set by the previous CAL:CONF:VOLT <range>,
ZERO | FSCale
command.

<ref_volts> must be within 4% of the actual reference voltage value as read after

CAL:CONF:VOLT or an error 3042 ’0x400: DSP-DAC adjustment went to
limit’ will be generated. If the reading on the external reference voltmeter is in
excess of 4% error from nominal voltage, verify the voltmeter and execute
*TST? on the VT1422A.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

ref_volts

numeric (float32)

must be within +10% or -50% of the

"expected" source output value

volts

Advertising
This manual is related to the following products: