Command sequence, Calibration:zero, Comments – VXI VT1422A User Manual

Page 277: Usage

Advertising
background image

VT1422A Command Reference 275

Chapter 7

CALibration

<ref_volts> may be specified in millivolts (mV).

When Accepted: Not while INITiated

Related Commands: CAL:CONF:VOLT, CAL:STORE ADC

Send with VXIplug&play Function: hpe1422_cmd(...)

Command

Sequence

CAL:CONF:VOLTAGE 4,FSCALE

*OPC?

Wait for operation to complete

enter statement

(now measure voltage with external DMM)

CAL:VAL:VOLT <measured value>

Send measured value to module

CALibration:ZERO?

CALibration:ZERO?

corrects Analog to Digital converter offset for any drift since

the last *CAL? or CAL:ZERO? command was executed. The offset calibration takes
about 5 seconds and should be done as often as the control set up allows.

Comments

The CAL:ZERO? command only corrects for A/D offset drift (zero). Use the

*CAL? common command to perform on-line calibration of channels as well as
A/D offset. *CAL? performs gain and offset correction of the A/D and each
channel with an analog SCP installed (both input and output).

Returned Value:

The data type for this returned value is int16.

Executing this command does not alter the module's programmed state

(function, range, etc.).

Related Commands: *CAL?

*RST Condition: A/D offset performed

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

Usage

CAL:ZERO?

enter statement here

returns 0 or -1

Value

Meaning

Further Action

0

Cal OK

None

-1

Cal Error

Query the Error Queue (SYST:ERR?)

See “Error Messages” on page 453