VT1422A Command Reference 277

Chapter 7

DIAGnostic

DIAGnostic:CALibration:SETup[:MODE]

DIAGnostic:CALibration:SETup[:MODE] <mode>

sets the type of calibration to

use for analog output SCPs like the VT1531A and VT1532A when *CAL? or
CAL:SET are executed.

Parameters

Comments

•

When <mode> is set to 1 (the *RST Default) channels are calibrated using the

Least Squares Fit method to provide the minimum error overall (over the entire
output range). When <mode> is 0, channels are calibrated to provide the
minimum error at their zero point. See the SCPs User's Manual for its accuracy
specifications using each mode.

•

Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?

•

*RST Condition: DIAG:CAL:SET:MODE 1

•

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

set analog DAC SCP cal mode for best zero accuracy

DIAG:CAL:SET:MODE 0

set mode for best zero cal

*CAL?

start channel calibration

DIAGnostic:CALibration:SETup[:MODE]?

DIAGnostic:CALibration:SETup[:MODE]?

returns the currently set calibration

mode for analog output DAC SCPs.

Comments

•

Returns a 1 when channels are calibrated using the Least Squares Fit method to

provide the minimum error overall (over the entire output range). Returns a 0
when channels are calibrated to provide the minimum error at their zero point.
See the SCPs User's Manual for its accuracy specifications using each mode.
The data type is int16.

•

Related Commands: DIAG:CAL:SET:MOD, *CAL?, CAL:SET

•

*RST Condition: DIAG:CAL:SET:MODE 1

•

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

mode

boolean (uint 16)

0 | 1

volts