282 VT1422A Command Reference

Chapter 7

DIAGnostic

DIAGnostic:CUSTom:REFerence:TEMPerature

DIAGnostic:CUSTom:REFerence:TEMPerature

extracts the current Reference

Temperature Register Contents, converts it to 32-bit floating point format and sends
it to the FIFO. This command is used to verify that the reference temperature is as
expected after measuring it using a custom reference temperature EU conversion table.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

the program must have EU table values stored in table_block

download the new reference EU table

DIAG:CUST:PIECEWISE <table_range>,<table_block>,(@<ch_list>)

designate channel as reference

SENS:FUNC:CUST:REF <range>,(@<ch_list>)

set up scan list sequence (ch 0 in this case)

Now run the algorithm that uses the custom reference conversion table

dump reference temp register to FIFO

DIAG:CUST:REF:TEMP

read the diagnostic reference temperature value

SENS:DATA:FIFO?

DIAGnostic:IEEE

DIAGnostic:IEEE <mode>

enables (1) or disables (0) IEEE-754 NAN (Not A

Number) and

±INF value outputs. This command was created for the VEE platform.

Parameters

Comments

When <mode> is set to 1, the module can return ±INF and NAN values according

to the IEEE-754 standard. When <mode> is set to 0, the module returns values
as ±9.9E37 for INF and 9.91E37 for NAN.

Related Commands: DIAG:IEEE?

*RST Condition: DIAG:IEEE 1

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

Set IEEE mode

DIAG:IEEE 1

INF values returned in IEEE standard

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

mode

boolean (uint 16)

0 | 1

volts