VT1422A Command Reference 289

Chapter 7

DIAGnostic

Test 5: This tests the filter settings on each bank (of eight channels).
The method of this test is to ensure that the approximate rise times
increase as the filters are changed from 100 Hz to 10 Hz and then to 2 Hz.
The list of possible error messages is shown below (NOTE that the 1xx
prefix to the channel number denotes the first three digits that uniquely
identify which VT1529A/B is to be tested – 100, 101, 108, 109, etc.):

1xx45 – Channel 0: 10 Hz rise time not at least 2x that of 100 Hz.
1xx46 – Channel 8: 10 Hz rise time not at least 2x that of 100 Hz.
1xx47 – Channel 16: 10 Hz rise time not at least 2x that of 100 Hz.
1xx48 – Channel 24: 10 Hz rise time not at least 2x that of 100 Hz.

1xx50 – Channel 0: 2 Hz rise time not at least 5x that of 100 Hz.
1xx51 – Channel 8: 2 Hz rise time not at least 5x that of 100 Hz.
1xx52 – Channel 16: 2 Hz rise time not at least 5x that of 100 Hz.
1xx53 – Channel 24: 2 Hz rise time not at least 5x that of 100 Hz.

1xx55 – Channel 0: 2 Hz rise time not at least 2x that of 10 Hz.
1xx56 – Channel 8: 2 Hz rise time not at least 2x that of 10 Hz.
1xx57 – Channel 16: 2 Hz rise time not at least 2x that of 10 Hz.
1xx58 – Channel 24: 2 Hz rise time not at least 2x that of 10 Hz.

The following errors are not likely to occur, but are possible:

1xx32 – Channel 0 100 Hz rise time test took too long.
1xx33 – Channel 8 100 Hz rise time test took too long.
1xx34 – Channel 16 100 Hz rise time test took too long.
1xx35 – Channel 24 100 Hz rise time test took too long.
1xx36 – Channel 0 10 Hz rise time test took too long.
1xx37 – Channel 8 10 Hz rise time test took too long.
1xx38 – Channel 16 10 Hz rise time test took too long.
1xx39 – Channel 24 10 Hz rise time test took too long.
1xx40 – Channel 0 2 Hz rise time test took too long.
1xx41 – Channel 8 2 Hz rise time test took too long.
1xx42 – Channel 16 2 Hz rise time test took too long.
1xx43 – Channel 24 2 Hz rise time test took too long.

Failure Information for -1 return. Probable causes:

a. Unable to communicate with VT1529A/B (is cable connected?)
b. Invalid channel number or multiple channels specified
c. Not enough memory to allocate internal arrays to hold data
d. VT1422A is currently performing a calibration operation.
e. VT1422A is currently performing a measurement operation.

Related Commands: *TST?, *CAL?, CAL:REMote?, SYST:ERR?

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

Usage

DIAG:REM:TEST? (@10000:10900)

self-test 4 RSCs at chs 00, 01, 08, and 09