Input:gain, Parameters, Comments – VXI VT1422A User Manual

Page 303: Usage, Input:low, Input:gain? input:low

Advertising
background image

VT1422A Command Reference 301

Chapter 7

INPut

INPut:GAIN?

INPut:GAIN? (@<channel>)

returns the gain currently set for channel. If the

channel is not on an input SCP, the query will return zero.

Parameters

Comments

The <channel> parameter must specify a single channel only.

If the channel specified does not have a programmable amplifier, INP:GAIN?

will return the nominal as-designed gain for that channel.

Returned Value: Numeric value as set by the INP:GAIN command.

The data type is float32.

When Accepted: Not while INITiated

Related Commands: INP:GAIN

*RST Condition: gain set to 1

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

Usage

INPUT:GAIN? (@105)

Check gain on channel 5

INP:GAIN? (@100)

Check gain on channel 0

INPut:LOW

INPut:LOW <wvolt_type>,(@<ch_list>)

controls the connection of input LO at a

Strain Bridge SCP channel specified by <ch_list>. LO can be connected to the
Wagner Voltage tap for quarter or half bridge configurations or disconnected for full
bridges. Note the VT1529A/B’s Wagner Voltage connection is only controlled by
the command “[SENSe:]STRain:BRIDge[:TYPE]” on page 362.

Parameters

Comments

Related Commands: INP:LOW?

*RST Condition: INP:LOW FLOAT (all Option 21 channels)

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

INP:LOW WVOL (@100:103,116:119)

connect LO of channels 0 through 3 and
16 through 19 to Wagner Ground.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

wvolt_type

discrete (string)

FLOat | WVOLtage

none

ch_list

channel list (string)

100 - 163

none