302 VT1422A Command Reference

Chapter 7

INPut

INPut:LOW?

INPut:LOW? (@<channel>)

returns the LO input configuration for the channel

specified by <channel>. This command is for strain SCPs only, not for VT1529A/B.

Parameters

Comments

The <channel> parameter must specify a single channel only.

Returned Value: Returns FLO or WV. The data type is string.

Related Commands: INP:LOW

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

INP:LOW? (@103)

enter statement will return either FLO or
WV for channel 3

INPut:POLarity

INPut:POLarity <mode>,<ch_list>

sets logical input polarity on a digital

SCP channel.

Parameters

Comments

If the channels specified are on an SCP that doesn't support this function, an

error will be generated. See the SCP's User's Manual to determine its capabilities.

Related Commands: for output sense; SOURce:PULSe:POLarity

*RST Condition: INP:POL NORM for all digital SCP channels.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

INP:POL INV,(@140:143)

invert first 4 channels on SCP at SCP
position 5. Channels 40 through 43

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

mode

discrete (string)

NORMal | INVerted

none

ch_list

string

100 - 163

none

Advertising