304 VT1422A Command Reference

Chapter 7

MEASure

MEASure

The MEASure subsystem provides convenient setup-and-execution for some
pre-measurement strain operations.

Subsystem Syntax

MEASure

:VOLTage

:EXCitation (@<ch_list>)
:UNSTrained (@<ch_list>)

MEASure:VOLTage:EXCitation?

MEASure:VOLTage:EXCitation? (@<ch_list>)

This command automatically

configures the VT1422A to measure the bridge excitation voltage at each channel in
<ch_list> and starts a measurement scan. It averages 32 measurements for each
channel and the averaged values are stored internally for later use by the strain
Engineering Unit Conversion process. The average of each channel’s reading is also
sent to the FIFO buffer for use in user defined conversion processes. The command
returns a single value which is the number of readings sent to the FIFO.

Note

The maximum excitation voltage the VT1422A can sense through the
VT1529A/B’s excitation sense path is 16 volts (±8 V dc centered about the Gnd
terminal). If a higher excitation voltage is supplied through the VT1529A/B, don’t
connect the excitation sense terminals.

Note that this command executes a measurement scan without executing any
algorithms that might be defined.

The sequence of individual commands to approximate this operation is

TRIGger:COUNt 1

one time through scan list

ROUTe:SEQuence:DEFine (@<ch_list>)

input the list of channels to measure

SENSe:FUNCtion:VOLTage [<range>,](@<ch_list>) set measurement function to volts

SENSe:STRain:EXCitation:STATE ON,(@<ch_list>) turn on excitation supplies

SENSe:STRain:CONNect EXCite,(@<ch_list>)

connect channel sense to excitation
supply

INIT

start measurement scan

SENSe:DATA:FIFO:COUNT?

query for number of readings in FIFO

enter statement here to return FIFO reading <count>

SENSe:DATA:FIFO:PART? <count>

read excitation values from the FIFO

enter statement to for block data from FIFO

next the excitation voltage values acquired above must be sent back to the VT1422A by
executing the following command once for each channel in <ch_list> above:

SENSe:STRain:EXCitation <voltage_value>,(@<channel>)