306 VT1422A Command Reference

Chapter 7

MEASure

MEASure:VOLTage:UNSTrained?

MEASure:VOLTage:UNSTrained? (@<ch_list>)

This command automatically

configures the VT1422A to measure the bridge output voltage at each channel in
<ch_list> and initiates a measurement scan. It averages 32 measurements for each
channel and the averaged values are stored for later use by the strain Engineering
Units conversion process on these channels. The strain bridges must be unstrained
during this time. The average of each channel’s reading is also sent to the FIFO buffer
to be viewed later if desired. The command returns a single value which is the number
of readings sent to the FIFO.

Note that this command executes a measurement scan without executing any
algorithms that might be defined.

The sequence of individual commands to approximate this operation is

TRIGger:COUNt 1

one time through scan list

ROUTe:SEQuence:DEFine (@<ch_list>)

input the list of channels to measure

SENSe:FUNCtion:VOLTage [<range>,](@<ch_list>) set measurement function to volts

SENSe:STRain:EXCitation:STATE ON,(@<ch_list>) turn on excitation supplies

SENSe:STRain:CONNect BRIDge,(@<ch_list>)

connect channel sense to bridge output

INIT

start measurement scan

SENSe:DATA:FIFO:COUNT?

query for number of readings in FIFO

enter statement here to return FIFO reading <count>

SENSe:DATA:FIFO:PART? <count>

read unstrained voltage readings from
the FIFO buffer

enter statement to for block data from FIFO

next the unstrained voltage values acquired above must be sent back to the VT1422A by
executing the following command once for each channel in <ch_list> above:

SENSe:STRain:UNSTrained <voltage_value>,(@channel)

Notes

1. Unlike the MEAS:VOLT:UNST? command, the individual command

sequence above cannot keep defined algorithms from running at INIT.
Since algorithms can place values into the FIFO buffer, it will be necessary to
determine which FIFO values are the excitation voltages.

2. Remember that the MEAS:VOLT:UNST? command also provides the

average of 32 measurements for each excitation value.

Parameters

Comments

This command is only for use on channels measured with the VT1529A/B. If

executed on channels connected to other strain SCPs, a 3007 "Invalid signal
conditioning plug-on" error message will be generated.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

ch_list

channel list (string)

10000 - 15731

none