314 VT1422A Command Reference

Chapter 7

OUTPut

Returned Value: Numeric value of amplitude set. The data type is float32.

Related Commands: OUTP:CURR:AMPL

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

Usage

OUTP:CURR:AMPLITUDE? (@163)

Check SCP current set for channel 63
(returns +3.0E-5 or +4.88E-4)

OUTPut:CURRent[:STATe]

OUTPut:CURRent[:STATe] <enable>,(@<ch_list>)

enables or disables current

source on channels specified in <ch_list>.

Parameters

Comments

OUTP:CURR:STAT does not affect a channel's amplitude setting. A channel

that has been disabled, when re-enabled sources the same current set by the
previous OUTP:CURR:AMPL command.

OUTP:CURR:STAT is most commonly used to turn off excitation current to

four-wire resistance (and resistance temperature device) circuits during
execution of CAL:TARE for those channels.

When Accepted: Not while INITiated

Related Commands: OUTP:CURR:AMPL, CAL:TARE

*RST Condition: OUTP:CURR OFF (all channels)

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

OUTP:CURR OFF,(@100,108)

turn off current source channels 0 and 8

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

enable

boolean (uint16)

1 | 0 | ON | OFF

none

ch_list

channel list (string)

100 - 163

none