Output:polarity, Parameters, Comments – VXI VT1422A User Manual

Page 318: Output:shunt, Usage, Output:polarity? output:shunt

Advertising
background image

316 VT1422A Command Reference

Chapter 7

OUTPut

OUTPut:POLarity?

OUTPut:POLarity? (@<channel>)

returns the polarity on the digital output channel

in <channel>.

Parameters

Comments

The <channel> parameter must specify a single channel.

Returned Value: returns one of NORM or INV. The type is string.

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

OUTPut:SHUNt

OUTPut:SHUNt <enable>,(@<ch_list>)

adds shunt resistance to one leg of bridge

on Strain Bridge Completion SCPs and the VT1529A/B Remote Strain Bridge unit.
This can be used for diagnostic purposes and characterization of bridge response.

Parameters

Comments

If <ch_list> specifies a non strain SCP, a 3007 "Invalid signal conditioning

plug-on" error is generated.

Only one channel on any one VT1529A/B can be specified in <ch_list>. This

is because a single resistor is used to shunt each of a VT1529A/Bs 32 channels.
The <ch_list> may specify one channel on each of several VT1529A/Bs.

When Accepted: Not while INITiated

Related Commands: [SENSe:]FUNCtion:STRain

…, [SENSe:]STRain…

*RST Condition: OUTP:SHUNT 0 on all Strain SCP and RSC channels

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

OUTP:SHUNT 1,(@116:119)

add shunt resistance at channels 16
through 19

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

string

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

enable

boolean (uint16)

0 | 1 | ON | OFF

none

ch_list

channel list (string)

100 - 15731

none