Usage, Output:shunt:source, Parameters – VXI VT1422A User Manual

Page 320: Comments

Advertising
background image

318 VT1422A Command Reference

Chapter 7

OUTPut

Related Commands: OUTPut:SHUNt

…, SENSe:FUNCtion:STRain…,

[SENSe:]STRain

*RST Condition: OUTP:SHUNT:SOURCE INT on all VT1529A/B channels

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

OUTP:SHUNT:SOUR EXT,(@10000,10800)

select user supplied shunt resistor on
VT1529As connected to channels 0, 1, 8,
and 9

OUTPut:SHUNt:SOURce?

OUTPut:SHUNt:SOURce? (@<channel>)

returns the source of the shunt

resistance on the specified VT1529A/B Strain channel.

Parameters

Comments

The <channel> parameter must specify a single channel only and since there is

a single shunt resistor for all channels on a VT1529A/B, it can be any channel
on the VT1529A/B.

If <channel> specifies a non VT1529A/B channel, a 3007 "Invalid signal

conditioning plug-on" error is generated.

Returned Value: Returns "INT" or "EXT." The data type is string.

Related Commands: OUTP:SHUNT:SOUR

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

OUTPUT:SHUNT:SOURCE? (@11600)

Check source of shunt resistance on
VT1529A/B connected to channel 16

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

10000 - 15731

none