VT1422A Command Reference 337

Chapter 7

[SENSe]

[SENSe:]DATA:FIFO:PART?

[SENSe:]DATA:FIFO:PART? <n_values>

returns <n_values> from the FIFO

buffer.

Parameters

Comments

Use the DATA:FIFO:COUNT? command to determine the number of values in

the FIFO buffer.

The format of values returned is set using the FORMat[:DATA] command.

Returned Value: ASCII values are returned in the form

±1.234567E±123.

For example 13.325 volts would be +1.3325000E+001. Each value is followed
by a comma (,). A line feed (LF) and End-Or-Identify (EOI) follow the last value.
The data type is a string array.

REAL 32, REAL 64 and PACK 64, values are returned in the IEEE-488.2-1987
Definite Length Arbitrary Block Data format. This data return format is
explained in “Arbitrary Block Program and Response Data” on page 233.
For REAL 32, each value is 4 bytes in length (the data type is a float32 array).
For REAL 64 and PACK 64, each value is 8 bytes in length (the data type is a
float64 array).

Note

Values which are a positive overvoltage return IEEE +INF and a negative overvoltage
return IEEE -INF (see Table 7-1 on page 294 for actual values for each data format).

Related Commands: DATA:FIFO:COUNT?

*RST Condition: FIFO buffer empty

Use VXIplug&play function: hpe1422_readFifoFast_Q(...)

Usage

DATA:FIFO:PART? 256

return 256 values from FIFO

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

n_values

numeric (int32)

1 - 2,147,483,647