VT1422A Command Reference 339

Chapter 7

[SENSe]

[SENSe:]FREQuency:APERture?

[SENSe:]FREQuency:APERture? <channel>

returns the frequency counting

gate time.

Parameters

Comments

If the channel specified is on an SCP that doesn't support this function, an error

will be generated. See the SCP's User's Manual for its capabilities.

Related Commands: SENSe:FREQuency:APERture

Returned Value: Returns numeric gate time in seconds. The type is float32.

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

[SENSe:]FUNCtion:CONDition

[SENSe:]FUNCtion:CONDition <ch_list>

sets the SENSe function to input the

digital state for channels in <ch_list>. Also configures digital SCP channels as inputs
(this is the *RST condition for all digital I/O channels).

Parameters

Comments

The VT1533A SCP senses 8 digital bits on each channel specified by this

command. The VT1534A SCP senses 1 digital bit on each channel specified by
this command.

If the channels specified are not on a digital SCP, an error will be generated.

Use the INPut:POLarity command to set input logical sense.

Related Commands: INPut:POLarity

*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital

SCP channels.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

To set second 8-bits of VT1533A at SCP position 4 and upper 4-bits of VT1534A at
SCP position 5 to digital inputs send:

SENS:FUNC:COND (@133,144:147)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

string

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

ch_list

string

100 - 163

none