VT1422A Command Reference 343

Chapter 7

[SENSe]

[SENSe:]FUNCtion:CUSTom:TCouple

[SENSe:]FUNCtion:CUSTom:TCouple <type>,[<range>,](@<ch_list>)

links channels with the custom Engineering Unit Conversion table loaded with the
DIAG:CUST:PIECE command. The table is assumed to be for a thermocouple and
the <type> parameter will specify the built-in compensation voltage table to be used
for reference junction temperature compensation. SENS:FUNC:CUST:TC allows an
EU table to be used that is custom matched to thermocouple wire that has been
characterized. Contact a VXI Technology System Engineer for more information on
Custom Piecewise Engineering Unit Conversion for specific applications.

Parameters

Comments

•

See “Creating and Loading Custom EU Conversion Tables” on page 150.

•

The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and

16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified that is larger than one of the first four
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the
16 V dc range). Specifying a value larger than 16 generates an error. Specifying
0 selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range.
The default range (no range parameter specified) is auto range.

•

If using amplifier SCPs, set them first and keep their settings in mind when

specifying a range setting. For instance, if the expected signal voltage is to be
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8,
<range> must be set no lower than 1 V dc or an input out-of-range condition
will exist.

•

The <type> EEXTended applies to E type thermocouples at 800

°C and above.

•

The *CAL? command calibrates temperature channels based on Sense Amplifier

SCP setup at the time of execution. If SCP settings are changed, those channels
are no longer calibrated. *CAL? must be executed again.

•

Related Commands: DIAG:CUST:PIEC, *CAL?,SENS:REF, and

SENS:REF:TEMP

•

*RST Condition: all custom EU tables erased

•

Send with VXIplug&play Function: hpe1422_cmd(...)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

type

discrete (string)

E | EEXT | J | K | N | R | S | T

none

range

numeric (float32)

see comments

V dc

ch_list

channel list (string)

100 - 163

none