VT1422A Command Reference 345

Chapter 7

[SENSe]

[SENSe:]FUNCtion:HVOLtage

[SENSe:]FUNCtion:HVOLtage [<range>,](@<ch_list>)

links the specified

VT1529B channels to return dc voltage on the high-level input (Excitation Sense +
and -) pins on the RJ-45 connector for the channel.

This command is valid only on the VT1529B and is available with VXIplug&play
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).

Parameters

Comments

The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and

16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified that is larger than one of the first four
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the
16 V dc range). Specifying a value larger than 16 causes an error. Specifying 0
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range.
The default range (no range parameter specified) is auto range.

The VT1529A/B has a fixed gain of 32. Keep this in mind when <range> is set.

See “DCV Measurement Command Sequence” on page 177.

When Accepted: Not while INITiated

Related Commands: CAL:REMote?

*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set

to FBEN on all strain channels.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

FUNC:HVOL (@10000:10001)

Channels 0 - 1 measure high-level input
voltage in auto-range (defaulted)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

range

numeric (float32)

see comments

V dc

ch_list

channel list (string)

10000 - 15731

none