Usage, Sense:]function:temperature, Parameters – VXI VT1422A User Manual

Page 353: Comments

Advertising
background image

VT1422A Command Reference 351

Chapter 7

[SENSe]

Related Commands: CAL:REMote?,[SENSE:]STRAIN

*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set

to FBEN on all strain channels

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:FUNC:STRAIN:QUAR:POST 1,(@10000:10007) quarter bridge conversion for VT1529B

channels 10000 to 10007

SENSE:FUNC:STRAIN:FBEN:POST 1,(@10800:10931) full bridge conversion for VT1529B

channels 10800 to 10931
(64 channels)

[SENSe:]FUNCtion:TEMPerature

[SENSe:]FUNCtion:TEMPerature <type>,<sub_type>,[<range>,](@<ch_list>)

links channels to an temperature EU conversion based on the sensor specified in
<type> and <sub_type>. Note: Do not include the thermocouple reference
temperature channels in this command (for that, use the SENS:REF
<type>,<sub_type>,(@<channel>) command).

Parameters

Comments

Resistance temperature measurements (RTDs and THERmistors) require the use

of Current Source Signal Conditioning Plug-Ons. The following table shows the
Current Source setting that must be used for the following RTDs and Thermistors:

The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and

16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified larger than one of the first four ranges,
the VT1422A selects the next higher range (for example, 4.1 selects the 16 V dc
range). Specifying a value larger than 16 generates an error. Specifying 0 selects
the lowest range (0.0625 V dc). Specifying AUTO selects auto range. The default
range (no range parameter specified) is auto range.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

type

discrete (string)

RTD | THERmistor | TCouple

none

sub_type

numeric (float32)
numeric (float32)

discrete (string)

for RTD, use 85 | 92
for THER, use 2250 | 5000 | 10000
for TC, use CUSTom | E | EEXT |
J | K | N | R | S | T

none

ohms

none

range

numeric (float32)

see comments

V dc

ch_list

channel list (string)

100 - 163

none

MAX (488 µA)

for RTD and THER,2250

MIN (30 µA)

for THER,5000 and THER,10000

Advertising
This manual is related to the following products: