356 VT1422A Command Reference

Chapter 7

[SENSe]

Parameters

Comments

•

See “Linking Input Channels to EU Conversion” on page 109

•

The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and

16 V dc. To select a range, simply specify the range value (for example, 4 selects
the 4 V dc range). If a value is specified that is larger than one of the first four
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the
16 V dc range). Specifying a value larger than 16 causes an error. Specifying 0
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range.
The default range (no range parameter specified) is auto range.

•

If using amplifier SCPs or RSCs, set them first and keep their settings in mind

when specifying a range setting. For instance, if the expected signal voltage is
to be approximately 0.1 V dc and the amplifier SCP for that channel has a gain
of 8, <range> must be set no lower than 1 V dc or an input out-of-range condition
will exist.

•

The <type> parameter specifies the sensor type that will be used to determine

the temperature of the isothermal reference panel. <type> CUSTom is
pre-defined as Type E with 0

°C reference junction temp and cannot be redefined.

•

For <type> THERmistor, the <sub_type> parameter may be specified in ohms

or kilohms.

•

The *CAL? command calibrates resistance channels based on Current Source

SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings
are changed, those channels are no longer calibrated. *CAL? must be executed
again.

•

Related Commands: SENSE:FUNC:TEMP

•

*RST Condition: Reference temperature is 0

°C.

•

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENSE:REF RTD,92,(@120)

sense the reference temperature on
channel 20 using an RTD

SENSE:REF THER,5000,(@113)

sense the reference temperature on
channel 113 using a 5k thermistor

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

type

discrete (string)

THERmistor | RTD | CUSTom

none

sub_type

numeric (float32)
numeric (float32)

for THER, use 5000

for RTD, use 85 | 92

for CUSTom, use 1

ohm

none
none

range

numeric (float32)

see comments

V dc

ch_list

channel list (string)

100 - 15731

none