Advertising
background image

VT1422A Command Reference 359

Chapter 7

[SENSe]

Parameters

Comments

Use SENS:FUNC:TEMP:POST to configure channels to measure thermocouples.

Then use SENS:REF:POST to configure one or more channels to measure an
isothermal reference temperature. Finally, use SENS:REF:CHAN:POST to
group the reference channel with its thermocouple measurement channels.

If thermocouple measurements are made through more than one isothermal

reference panel, set up a reference channel for each. Execute the
SENS:REF:CHAN:POST command for each reference/measurement channel
group.

Related commands: SENS:FUNC:TEMP:POST, SENS:REF:POST

*RST Condition: No channels are linked to references.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:FUNC:TEMP:POST TC,J,.0625,(@10008:10015) J type TCs on channels 8 through 15

SENS:REF:POST THER,5000,4,(@10000),(@10001) sense the reference temperature on

ch 10001; excitation on ch 10000;
measure on 4 V range

SENS:REF:CHAN (@10001),(@10008:10015)

link TC channels to thermistor

ROUT:SEQ:DEF (@10008:10015,10001,10000)

set up scan list

[SENSe:]REFerence:TEMPerature

[SENSe:]REFerence:TEMPerature <degrees_c>

stores a fixed reference junction

temperature in the Reference Temperature Register. Use when the thermocouple
reference junction is kept at a controlled temperature.

Note

This reference temperature is used to compensate all subsequent thermocouple
measurements until the register is overwritten by another SENSE:REF:TEMP value
or by scanning a channel linked with the SENSE:REFERENCE command. If used,
SENS:REF:TEMP must be executed before scanning any thermocouple channels.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

ref_channel

channel list (string)

10000 - 15731

none

ch_list

channel list (string)

10000 - 15731

none