366 VT1422A Command Reference

Chapter 7

[SENSe]

[SENSe:]STRain:EXCitation:STATe

[SENSe:]STRain:EXCitation:STATe <enable>,(@<ch_list>)

connects or

disconnects all four excitation supply ports on a VT1529A/B. Only one channel for
each VT1529A/B needs to be specified in <ch_list> and all four excitation supply
ports on that unit will configure as specified in <enable>. The first channel number
on each possible VT1529A/B is: 10000, 10100, 10800, 10900, 11600, 11700, 12400,
12500, 13200, 13300, 14000, 14100, 14800, 14900, 15600, 15700.

Parameters

Comments

Related Commands: SENSE:STRAIN:

…, SENSE:FUNC:STRAIN…

*RST Condition: OFF

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

STRAIN:EXC:STAT ON,(@10800)

turn on all excitation supplies on
VT1529A/B connected to on-board
channel 08

[SENSe:]STRain:EXCitation:STATe?

[SENSe:]STRain:EXCitation:STATe? (@<channel>)

returns the state of all four

VT1529A/B excitation supply ports referenced in <channel>. The first channel
number on each possible VT1529A/B is: 10000, 10100, 10800, 10900, 11600, 11700,
12400, 12500, 13200, 13300, 14000, 14100, 14800, 14900, 15600, 15700.

Parameters

Comments

Related Commands: SENSE:STRAIN:EXC:STAT

Returned Value: Numeric, 0 or 1. Type is uint16.

*RST Condition: OFF

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

Usage

STRAIN:EXC:STAT? (@10800)

check state of excitation supply ports on
VT1529A/B connected to channel 08

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

enable

boolean (uint16)

ON | OFF

none

ch_list

channel list (string)

10000 - 15731

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

1000 - 15731

none