Usage, Sense:]totalize:reset:mode, Parameters – VXI VT1422A User Manual

Page 372: Comments

Advertising
background image

370 VT1422A Command Reference

Chapter 7

[SENSe]

Related Commands: STRAIN:UNST

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

Usage

STRAIN:UNST? (@107)

query unstrained voltage for channel 7

enter statement here

returns the unstrained voltage set by
STR:UNST

[SENSe:]TOTalize:RESet:MODE

[SENSe:]TOTalize:RESet:MODE <select>,<ch_list>

sets the mode for resetting

totalizer channels in <ch_list>.

Parameters

Comments

In the INIT mode the total is reset only when the INITiate command is executed.

In the TRIGger mode the total is reset every time a new scan is triggered.

If the channels specified are not on a Frequency/Totalize SCP, an error will

be generated.

Related Commands: SENS:FUNC:TOT, INPUT:POLARITY

*RST Condition: SENS:TOT:RESET:MODE INIT

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SENS:TOT:RESET:MODE TRIG,(@134)

totalizer at channel 34 resets at each
trigger event

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

select

discrete (string)

INIT | TRIGger

seconds

ch_list

string

100 - 163

none

Advertising