Comments, Source:pulm:state, Parameters – VXI VT1422A User Manual

Page 377: Source:pulse:period, Source:pulm:state? source:pulse:period

Advertising
background image

VT1422A Command Reference 375

Chapter 7

SOURce

Comments

This command is coupled with the SOURce:FM command. If the FM state is

enabled then the PULM state is disabled. If the PULM state is enabled then the
FM state is disabled. If both the FM and the PULM states are disabled then the
PULSe channel is in the single pulse mode.

If the channels specified are not on a Frequency/Totalize SCP, an error will

be generated.

*RST Condition: SOUR:PULM:STATE OFF

Send with VXIplug&play Function: hpe1422_cmd(...)

SOURce:PULM:STATe?

SOURce:PULM[:STATe]? (@<channel>)

returns the pulse width modulated mode

state for the PULSe channel in <channel>.

Parameters

Comments

The <channel> parameter must specify a single channel.

Returned Value: returns 1 (on) or 0 (off). The type is int16.

Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)

SOURce:PULSe:PERiod

SOURce:PULSe:PERiod <period>,(@<ch_list>)

sets the fixed pulse period value

on a pulse width modulated pulse channel. This sets the frequency (1/period) of the
pulse-width-modulated pulse train.

Parameters

Comments

If the channels specified are not on a Frequency/Totalize SCP, an error will

be generated.

*RST Condition: SOUR:FM:STATE OFF and SOUR:PULM:STATE OFF

Related Commands: SOUR:PULM:STATE, SOUR:PULS:POLarity

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

string

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

period

numeric (float32)

25E-6 to 7.8125E-3

(resolution 0.238

μs)

seconds

ch_list

string

100 - 163

none

Advertising