VT1422A Command Reference 377

Chapter 7

SOURce

The variable frequency control for this channel is provided by the algorithm

language. When the algorithm executes an assignment statement to this channel,
the value assigned will be the frequency setting. For example:

O143 = 2000 /* set channel 43 to 2 kHz */

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SOUR:PULS:WIDTH 2.50E-3,(@143)

set fixed pulse width of 2.5 ms on
channel 43

SOURce:PULSe:WIDTh?

SOURce:PULSe:WIDTh? (@<ch_list>)

returns the fixed pulse width value on a

frequency modulated pulse channel.

Parameters

Comments

The <channel> parameter must specify a single channel.

If the channels specified are not on a Frequency/Totalize SCP, an error will

be generated.

Returned Value: returns the numeric pulse width. The type is float32.

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

SOURce:VOLTage[:AMPLitude]

SOURce:VOLTage[:AMPLitude] <-offset_v>,(@<ch_list>)

can be used to

reduce bridge offset voltage present at the dynamic strain "Buffered Output"
channel connectors.

Parameters

Comments

If the channels specified are not on a VT1529A/B, an error will be generated.

*RST Condition: SOUR:VOLT 0,(@ <all VT1529A/B channels>)

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

string

100 - 163

none

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

-offset_v

numeric (float32)

-1.651 to +1.664 | MIN | MAX

(resolution 13 mV)

volts

ch_list

string

10000 - 15731

none