378 VT1422A Command Reference

Chapter 7

SOURce

To reduce the offset voltage at each dynamic strain "Buffered Output" channel:

1. Measure a Buffered Output channel with its bridge unstrained and place

the value in a variable arbitrarily called <offset_v>.

2. Send minus <offset_v> to that channel with the SOUR:VOLT command.

For example: SOUR:VOLT <-offset_v>,(@10000)

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SOUR:VOLT .041,(@10031)

correct a -41 mV offset at channel 31 of
VT1529A/B 00.