392 VT1422A Command Reference

Chapter 7

SYSTem

Parameters

Comments

The <channel> parameter must specify a single channel only. Either the 3-digit

or 5-digit format can be used.

Returned Value: Examples of the response string format are:

E1539 Strain SCP; Chan 108 = E1529A; Chan 109 = E1529B
E1539 Strain SCP; Chan 108 = No Device; Chan 109 = E1529B

The data type is string.

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

SYST:CTYPE? (@100)

return SCP type installed at channel 0

SYSTem:ERRor?

SYSTem:ERRor?

returns the latest error entered into the Error Queue.

Comments

SYST:ERR? returns one error message from the Error Queue (returned error is

removed from queue). To return all errors in the queue, repeatedly execute
SYST:ERR? until the error message string = +0, "No error"

Returned Value: Errors are returned in the form:

±<error number>, "<error message string>"

RST Condition: Error Queue is empty.

Use VXIplug&play function: hpe1422_error_query(...)

Usage

SYST:ERR?

returns the next error message from the
Error Queue

SYSTem:VERSion?

SYSTem:VERSion?

returns the version of SCPI this instrument complies with.

Comments

Returned Value: String "1990." The data type is string.

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

SYST:VER?

Returns "1990"