Usage, Trigger:source, Trigger:timer[:period – VXI VT1422A User Manual

Page 399: Parameters, Comments, Trigger:source? trigger:timer[:period

Advertising
background image

VT1422A Command Reference 397

Chapter 7

TRIGger

While TRIG:SOUR is IMM, simply INITiate the trigger system to start a

measurement scan.

When Accepted: Before INIT only.

Related Commands: ABORt, INITiate, *TRG

*RST Condition: TRIG:SOUR TIMER

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

TRIG:SOUR EXT

Hardware trigger input at Connector
Module

TRIGger:SOURce?

TRIGger:SOURce?

returns the current trigger source configuration.

Returned Value: Discrete; one of BUS, EXT, HOLD, IMM, SCP, TIM, or

TTLT0 through TTLT7. The data type is string. See the TRIG:SOUR command
for more response data information.

Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

Usage

TRIG:SOUR?

ask VT1422A to return trigger source
configuration

TRIGger:TIMer[:PERiod]

TRIGger:TIMer[:PERiod] <trig_interval>

sets the interval between scan triggers.

Used with the TRIG:SOUR TIMER trigger mode.

Parameters

Comments

In order for the TRIG:TIMER to start it must be Armed. For information on

timer arming see the ARM subsystem in this command reference.

The default interval is 10E-3 seconds. interval may be specified in seconds,

milliseconds (ms) or microseconds (us). For example; 0.0016, 1.6 ms or
1600 us. The resolution for <interval> is 100 µs.

TRIG:TIMer periods shorter than the value returned from the

ALG[:EXPL]:TIME? command may result in "Trigger too fast" errors.

When Accepted: Before INIT only.

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

trig_interval

numeric (float32)

(string)

100E-6 to 6.5536 |

MIN | MAX

seconds