78 Programming the VT1422A & VT1529A/B for Remote Strain Measurement

Chapter 3

Programming for Remote Strain Measurement

This programming section is focused exclusively on programming the
VT1422A and VT1529A/B for remote strain measurement. For more
general VT1422A programing see Chapter 4 “Programming the VT1422A
for Data Acquisition and Control”.

Power-on and *RST

Configuration

Some of the programming operations that follow may already be set after
Power-on or after an *RST command. Where these default settings coincide
with the configuration settings required, there is no need to execute a
command to set them. These are the default settings:

No channels defined in scan list.

Programmable SCPs configured to their Power-on defaults.

VT1529A/B input filters:

-- INPut:FILTer:FREQuency 10,(@<all channels>)
-- INPut:FILTer:STATe ON,(@<all channels>)

All analog input channels linked to the EU conversion for voltage.

ARM:SOURce

IMMediate

TRIGger:SOURce

TIMer

TRIGger:COUNt

1

TRIGer:TIMer

.010

(10 ms)

FORMat

ASC,7

(ASCII)

SENSe:DATA:FIFO:MODE

BLOCking

The defaults for the STRain Subsystem when SENS:FUNC:STRain is

selected will be:
-- Unstrained voltage for all strain channels is assumed to be zero.
-- Gage factor for all strain channels is assumed to be 2.
-- Excitation voltage for all strain channels is assumed to be 1.0E6

(must be changed to actual value to make reasonable measurements).

The default for the VT1529A/B strain configuration switches is:

-- Full Bridge (FBEN) on all 32 Channels (SENS:STR:BRID FBEN).
-- Bridge output sensed.

Description of

Strain Measurement

This section describes the three ways to make strain measurements with the
VT1529A/B. It includes references to SCPI commands as well as command
sequences to perform the strain measurements described.

CALibration First

To make proper measurements, *CAL? and CAL:REMote? should have
been done first. Perform *CAL? and CAL:REMote? before making
important measurement runs or if the temperature of the instrument’s
environment has changed significantly. Remember, the accuracy
specifications given in Appendix A on page 419 depend on recent *CAL?
and CAL:REMote? operations.