88 Programming the VT1422A & VT1529A/B for Remote Strain Measurement

Chapter 3

Voltage Conversion

Command Sequence

Below is an example VXIplug&play command sequence. Note that this is
not executable; it has been simplified for easier reading. The C++ example
source file (voltseq.cpp) is on the CD supplied with the instrument.
View the readme.txt file provided with the VXIplug&play driver for
example program file location.

Figure 3-18. Converting Bridge Voltage Measurements to Strain

Readings returned in voltage

Retrieve readings from

FIFO and/or CVT

Trigger event for each pass

through Scan List

Set up Trigger System to scan

strain bridge channels

Set up the analog input Scan List

Measure unstrained bridge output

voltages and retrieve from FIFO

Measure bridge excitation

voltages and retrieve from FIFO

Enable excitation voltage

to the strain bridges

Set bridge configuration switches

SENSe:STRain:BRIDge[:TYPE]

SENSe:STRain:EXCitation:STATe

MEASure:VOLTage:EXCitation?
SENS:DATA:FIFO:PART?

MEASure:VOLTage:UNSTrained?
SENSE:DATA:FIFO:PART?

ROUTe:SEQuence:DEFine

TRIG:SOURce, TRIG:COUNt, ARM:SOURce
TRIG:TIMer

Set up Sample Timer for best

channel-to-channel scan rate

Initiate Trigger System

SAMPle:TIMer

INITiate[:IMMediate]

TRIG[:IMM], *TRG (if TRIG:SOUR HOLD),
EXTernal trig, TTLTRGn signal, or TRIG:TIMer

SENSe:DATA:FIFO:..., SENSe:DATA:CVTable?

Set Input Filter on VT1529A/B

INPut:FILTer:FREQuency and INPut:FILTer:STATe

Post-process unstrained, excitation,

gage factor, and strained values

to calculate strain

SENSe:FUNCtion:VOLTage