Calibration and Troubleshooting

6-12

975-0200-01-01

SYST:COMM:APR:LEV:CURR 4

4. Set the input source as Resistive Non-Isolated for Current

Programming. The SCPI command is:

SYST:REM:SOUR:CURR ARES

5. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

6. Enter the interactive calibration mode by sending the SCPI command:

CAL:OUTP:ANAL:CURR:RES

7. Press the terminal keys – or + to reduce or increase the main output

current value. Adjust the main output current to 100%.

8. Press any key other than the + or – keys to exit from interactive

calibration mode.