Ohaus I150-S Indicator User Manual
Page 39
39
7.
Apply the required load to the load
receiver, and then press
ENTER
.
8.
The display will show “buSY” while
the I150 reads the last calibration
point.
9.
“CALib” will be displayed when cali-
bration is completed and the calibra-
tion submenu will be returned.
If two load receivers are being used,
you can press
DOWN
to switch to the
second load receiver, and repeat the
procedure.
NOTE: If calibration is aborted at any
time, previous calibration data will re-
main.
Calibration can be aborted by pressing
CLRR
or
OFFF
.
End
End
End
End
End
You must select “End” to exit the calibra-
tion submenu.
1.
Press
ENTR
when “End” is displayed.
2.
“CAL” will be displayed and the main
menu will be returned.
NOTE: Calibration data is permanently
stored in memory when you use “End” to
exit from the Main menu, or when the OFF
key is pressed.