Installation, Power connections – Ohaus I150-S Indicator User Manual

Page 7

Advertising
background image

7

INSTALLATION

INSTALLATION

INSTALLATION

INSTALLATION

INSTALLATION

Power Connections

Power Connections

Power Connections

Power Connections

Power Connections

Using a small screwdriver, connect the
power cord to TB2 located on the main
circuit board inside the enclosure as fol-
lows (see diagram):

1.

Pass the power cord through the seal
bushing of the liquid tight connector
on the lower right of the enclosure.

Load

Receiver #2

Cable

Load

Receiver #1

Cable

RS232

Cable

Power

Cord

TB1

TB3

TB2

2.

Connect the ground wire (green or
green/yellow) to the case ground
screw located inside the enclosure,
just above the seal bushing. Secure
the wire to the screw using the #6-32
hex nut provided in the accessory kit.

Tighten nut to minimum torque of 10
inch-pounds to ensure ground conti-
nuity.

3.

Connect the NEUTRAL wire (WHITE
or BLUE) to the terminal labeled NEU-
TRAL.

4.

Connect the BLACK or BROWN wire
to the terminal labeled with the ap-
propriate line voltage (100V, 120V,
220V or 240V).

CAUTION: Indicator may be dam-
aged if operated at incorrect voltage.

5.

Permanently mark the appropriate
line voltage box on the serial number
label located on the side of the enclo-
sure.

Case

Ground

Screw

Power

Cord

Advertising