Lab 960

Operation

25

ba75557e05

11/2009

4.3.5

Measuring with stability control

The stability control function (SC) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.

Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.

1

With <SC>, activate the stability control function.
The AR function display indicator appears.
The current measured value is frozen (hold function).

2

Start measurement with stability control with <OK>.
An AutoRead measurement is carried out to control the stability
of the measured value. The AR display indicator flashes until a
stable value is measured. The calibration process is finished
when AR stops flashing.
This measured value is downloaded to the interface.

3

If necessary, start the next measurement with stability control
with <OK>.

4

Press <MODE> or <SC>

to terminate the stability control.