5 measuring with stability control, Measuring with stability control – Xylem Lab 960 User Manual
Page 25

Lab 960
Operation
25
ba75557e05
11/2009
4.3.5
Measuring with stability control
The stability control function (SC) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.
Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.
1
With <SC>, activate the stability control function.
The AR function display indicator appears.
The current measured value is frozen (hold function).
2
Start measurement with stability control with <OK>.
An AutoRead measurement is carried out to control the stability
of the measured value. The AR display indicator flashes until a
stable value is measured. The calibration process is finished
when AR stops flashing.
This measured value is downloaded to the interface.
3
If necessary, start the next measurement with stability control
with <OK>.
4
Press <MODE> or <SC>
to terminate the stability control.