What to do if...
61
ba75557e05
11/2009
6
What to do if...
Error message
E1
Error message,
E3
Display of Cln
Display,
LoBat
Display
to
Instrument does not
react to keystroke
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Unsuitable calibration
solution
– Check calibration solutions
Cause
Remedy
– Cleaning interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press and hold the <SC>
key and switch the meter on