Lab 960

Technical data

65

ba75557e05

11/2009

Stand October 30, 2009

Translation of the legally binding German version

7.2

Measuring ranges, resolution, accuracy

Measuring ranges,

resolution

Variable

Measuring range

Resolution

ӕ

[µS/cm]

0.000 ... 1.999*
0.00 ... 19.99**
0.0 ... 199.9
0 ... 1999

0.001
0.01
0.1
1

ӕ

[mS/cm]

0.00 ... 19.99
0.0 ... 199.9
0 ... 500

0.01
0.1
1

Resistivity
[MOhm*cm]

0.00 ... 19.99
0.0 ... 199.9
0 ... 1999

0.01
0.1
1

SAL

0.0 ... 70.0
according to the IOT
table

0.1

TDS [mg/l]

0 ... 1999
Factor can be set
between 0.40 and
1.00

1

T [°C]

− 5.0 ... + 120.0

0.1

T [°F]

+ 23.0 ... + 248.0

0.1

* only possible with cells of the cell constant, 0.010 cm

-1

** only possible with cells of the cell constant, 0.010 cm

-1

or 0.090 ... 0.110 cm

-1

Cell constants

Cell constant C

Values

Can be calibrated in
the ranges

0.450 ... 0.500 cm

-1

0.585 ... 0.715 cm

-1

0.800 ... 1.200 cm

-1

Adjustable

0.250 ... 2.500 cm

-1

0.090 ... 0.110 cm

-1

0.010 cm

-1

(fixed)

Reference temperature

Reference
temperature

Values

Adjustable

20 °C (Tr20)
25 °C (Tr25)