2 automatic storage at intervals, Automatic storage at intervals – Xylem Lab 960 User Manual
Page 40
Advertising

Operation
Lab 960
40
ba75557e05
11/2009
4.6.2
Automatic storage at intervals
In order to store measured values automatically at certain time
intervals, set the storage interval (Int.1).
Setting the save interval
The default setting for the storage interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
Note
When the AutoStore function is active, the automatic switchoff function
is off.
1
Using <STO
_
>, call up the setting menu for automatic storing
(Int.1) .
2
Set the required interval with <
▲><▼>.
3
Confirm with <OK>.
FrEE, the number of free memory locations is displayed.
E
E
r
199